System maintenance
LOGIQ V3/V5/V5 Expert – Basic Service Manual                 10-17
5726264-100 English Rev.8
10-6-5 Probe maintenance
10-6-5-1 Probe related checks
10-6-5-2 Basic probe care
The Ultrasound system user manuals and various probe 
handling cards provide a complete description of probe care, 
maintenance, cleaning and disinfection. Ensure that you are 
completely familiar with the proper care of GE probes.
Ultrasound probes can be easily damaged by improper 
handling. See the User Manual and probe care cards for more 
details. Failure to follow these precautions can result in serious 
injury and equipment damage. Failure to properly handle or 
maintain a probe may also void its warranty.
Any evidence of wear indicates the probe cannot be used.
Do a visual check of the probe pins and Ultrasound system 
sockets before plugging in a probe.
The Interoperative probes often have special considerations and 
individual probe user manuals. For Interoperative probes also 
refer to their separate user manuals.
Table 10-11:  System preliminary checks
Step Item Description
1.  Probe Holder Clean probe holders. (they may need to be soaked to remove excess gel).
2.  Probes Thoroughly check the Ultrasound system probe connectors and remove 
dust from inside the connector sockets if necessary. Visually check for 
bent, damaged or missing pins.
3.  Probes Verify that the Ultrasound system properly recognizes all probes.