General Ultrasound system requirements
LOGIQ V3/V5/V5 Expert – Basic Service Manual 2-11
5726264-100 English Rev.8
2-2-5 Probes environmental requirements
2-2-5-1 Operation, storage and transport temperatures for probes
Probes should be operated, stored, or transported within the
parameters outlined below.
NOTE: Refer to Table 2-1 on page 2-3 to determine the needed
settlement time.
2-2-6 Time and manpower requirements
Site preparation takes time. Begin site preparation checks as
soon as possible, if possible, six weeks before delivery, to allow
enough time to make any changes.
CAUTION
Ensure that the probe face temperature does not exceed the
normal operation temperature range.
Table 2-6: Probe Environmental Requirements
Operational Storage Transport
Temperature 3° - 40° C
38° - 104° F
-5° - 50° C
23° - 122° F
-5° - 50° C
23° - 122° F
Humidity 30- 80%
non-condensing
10 - 90%
non-condensing
10 - 90%
non-condensing
Pressure 700 - 1060hPa 700 - 1060hPa 700 - 1060hPa
CAUTION
Check the room temperature before you use the probe.
CAUTION
Ensure that the probe face temperature does not exceed the
normal operation temperature range.

Table of Contents

Other manuals for GE LOGIQ V5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V5 and is the answer not in the manual?

GE LOGIQ V5 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V5
CategoryMedical Equipment
LanguageEnglish

Related product manuals