Electrical safety tests
LOGIQ V3/V5/V5 Expert – Basic Service Manual                 10-19
5726264-100 English Rev.8
10-7 Electrical safety tests
10-7-1 Safety test overview
The electrical safety tests in this section are based on and 
conform to IEC 60601-1 Medical Equipment Safety Standards. 
They are intended for the electrical safety evaluation of 
cord-connected, electrically operated, patient care equipment. If 
additional information is needed, refer to the IEC 60601-1 
documents
THE USER MUST ENSURE THAT THE SAFETY 
INSPECTIONS ARE PERFORMED AT LEAST EVERY 
12 MONTHS ACCORDING TO THE HISTORICAL 
DATA. ONLY TRAINED PERSONS ARE ALLOWED TO 
PERFORM THE SAFETY INSPECTIONS MENTIONED 
ABOVE.
DANGER TO MINIMIZE RISK OF ELECTRICAL SHOCK, ONLY 
TRAINED PERSONS ARE ALLOWED TO PERFORM THE 
ELECTRICAL SAFETY INSPECTIONS AND TESTS.
DANGER TO AVOID ELECTRICAL SHOCK, THE ULTRASOUND 
SYSTEM UNDER TEST MUST NOT BE CONNECTED TO 
OTHER ELECTRICAL EQUIPMENT. REMOVE ALL 
INTERCONNECTING CABLES AND WIRES. THE 
ULTRASOUND SYSTEM UNDER TEST MUST NOT BE 
CONTACTED BY USERS OR PATIENTS WHILE 
PERFORMING THESE TESTS.
Possible risk of infection. Do not handle soiled or contaminated 
probes and other components that have been in patient 
contact. Follow appropriate cleaning and disinfecting 
procedures before handling the equipment.