Replacement Procedures
8-8       LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
8-3-7 Loading the System Software
NOTE:  Before loading the system software, please ensure that the 
power can be continuously supplied and there is no risk of 
power cut off during loading procedure.
NOTE:  Before loading/reload the system software, please change 
system language to English. Otherwise, there will be garbage on 
the system after software loading/reload.
There is one method to load the system software:
•  Load the system software with USB memory stick.
8-3-7-1 Loading the C Disk with USB memory stick
NOTE:  While it is believed to be unnecessary, it would not hurt to 
disconnect the system from the network and remove all 
transducer.
NOTE:  Please ensure AC adapter is connected during system upgrade!
1.  Insert the USB memory stick labeled “System & Application 
Software” to the system.
2.  Properly turn off the scanner by momentarily pressing the 
Power On/Off Switch. In System-Exit window, select 
Shutdown to shutdown the system.
NOTE:  If the system will not shutdown normally, hold down the 
Power On/Off Switch until the light turns off.
3.  Power on the system. The system will detect the USB 
memory stick automatically.
While the software install procedure is designed to 
preserved data, you should save any patient data, 
images, system setups and customer presets to CD, 
DVD, USB Flash Drive, or USB Hard Disk before doing a 
software upgrade.