EMC, EMI and ESD
LOGIQ V3/V5/V5 Expert – Basic Service Manual             1-27
5726264-100 English Rev.8
1-10 EMC, EMI and ESD
1-10-1 Electromagnetic Compatibility (EMC)
Electromagnetic compatibility describes a level of performance 
of a device within its electromagnetic environment. This 
environment consists of the device itself and its surroundings 
including other equipment, power sources and persons with 
which the device must interface. Inadequate compatibility results 
when a susceptible device fails to perform as intended due 
interference from its environment or when the device produces 
unacceptable levels of emission to its environment. This 
interference is often referred to as radio–frequency or 
electromagnetic interference (RFI/EMI) and can be radiated 
through space or conducted over interconnecting power of 
signal cables. In addition to electromagnetic energy, EMC also 
includes possible effects from electrical fields, magnetic fields, 
electrostatic discharge and disturbances in the electrical power 
supply.
1-10-2 CE Compliance 
LOGIQ V3/V5/V5 Expert conforms to all applicable conducted 
and radiated emission limits and to immunity from electrostatic 
discharge, radiated and conducted RF fields, magnetic fields 
and power line transient requirements.
For applicable standards, refer to the Safety Chapter of the 
Ultrasound system User’s Manual.
NOTE:  For CE Compliance, it is critical that all covers, screws, 
shielding, gaskets, mesh, clamps, are in good condition, 
installed tightly without skew or stress. Proper installation 
following all comments noted in this service manual is required 