System Setup
3-20       LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
3-5-7 Connections on the I/O Rear Panel 
NOTE:  Accessory equipment connected to the analog and digital 
interfaces must be certified according to the respective IEC 
standards (e.g. IEC60950 for data processing equipment and 
IEC60601-1 for medical equipment). Furthermore, all complete 
configurations shall comply with the valid version of the system 
standard IEC60601-1. Everybody who connects additional 
equipment to the signal input part or signal output part of LOGIQ 
V3/V5/V5 Expert, configures a medical system, and is therefore 
responsible that the Ultrasound system complies with the 
requirements of the valid version of IEC60601-1. If in doubt, 
consult the technical service department or your local 
representative for GE.
3-5-7-1 Connect Ethernet
Connect the network cable to the Ethernet connector on the I/O 
Rear Panel.
The connector is located on the rear side of LOGIQ V3/V5/V5 
Expert.
3-5-7-2 Connect USB Flash Drive 
NOTE:  USB Flash Drive approved for LOGIQ V3/V5/V5 Expert are 
verified for EMC performance according to EN55011 class B. 
The use of any other USB Flash Drive will compromise this 
verification, and may cause interference on LOGIQ V3/V5/V5 
Expert itself, or on other electronic devices. 
For approved models, please refer to Chapter 9.
Insert the USB Flash Drive in one of the USB ports on the 
LOGIQ V3/V5/V5 Expert. 
3-5-8 Connecting probes
Please refer to Basic User Manual on how to connect/
disconnect a probe.
3-5-9 Powering the system
Please refer to Basic User Manual on how to power the system.