Care and Maintenance
10-22       LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
10-7-2 Leakage current limits(continued)
NOTE:  *Mains Applied refers to the sink leakage test where mains 
(supply) voltage is applied to the part to determine the amount of 
current that will pass (or sink) to ground if a patient contacted 
mains voltage.
The following tests are performed at the factory and should be 
performed at the site. These tests are: chassis leakage current, 
and probe leakage current. All measurements are made with an 
electrical safety analyzer which should be calibrated and 
compliant with AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551.
Table 10-14:  Type CF Applied Part Leakage Current Limits - ECG Connections
Country
Normal 
Condition
Open 
Ground
Reverse 
Polarity
Open 
Neutral
*Mains 
Applied
All 0.1 mA 0.5 mA 0.5 mA 0.5 mA 5.0 mA
Table 10-15:  Equipment Type and Test Definitions 
Applied 
Parts 
(AP)
Parts or accessories that contact the patient to 
perform their function. For ultrasound equipment, 
this includes transducers and ECG leads.
Type BF Body Floating or 
non-conductive ultrasound 
probes which are marked 
with the 'man in box' BF 
symbol. this includes all 
transducers.
Type CF Cardiac Floating or 
non-conductive 
intraoperative probes for 
direct cardiac contact and 
isolated ECG connections 
so marked with the 'heart in 
box' CF symbol.
Sink 
Leakage
The current resulting from the application of mains 
voltage to the applied part. This test is required 
test for Type CF applied parts.