Electrical safety tests
LOGIQ V3/V5/V5 Expert – Basic Service Manual                 10-29
5726264-100 English Rev.8
10-7-6-6 Data Sheet for Transducer Source Leakage Current
The test passes when all readings measure less than the values 
shown in Table 10-13 on page 10-21. Record all data on the PM 
Inspection Certificate.
Equipment damage possibility. Never switch the Polarity and 
the status of Neutral when the Ultrasound system is powered 
ON. Be sure to turn the Ultrasound system power OFF before 
switching them using the POLARITY switch and/or the 
NEUTRAL switch. Otherwise, the Ultrasound system may be 
damaged.
Table 10-17:  Typical Data Sheet For Transducer Source Leakage Current
Transducer Tested:
 Unit Power
Tester Power Polarity 
Switch
Tester GROUND or 
NUETRAL Switch
Measurement
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
OFF NORM OPEN
OFF NORM CLOSED
OFF REV OPEN
OFF REV CLOSED