Maintenance
4.Snapthebackpanelintoplaceinthebottomcover.
5.ReconnectallribbonsandcablestothePWA.
6.Replacethetopcoverassembly.
See“ReplacingtheTopCover”onpage79forinstructions.
7.Reinsertthebattery.
Seestep6throughstep8in“ReplacingtheBattery”onpage74forinstructions.
8.Turnonthesystem.
9.UsethefollowingproceduretorestoresystemsetupsthatweresavedtotheSD
cardbeforethenewPWAwasinstalled:
a.InserttheSDcardwiththesavedsystemsetupsintotheSDcardslot.
b.OntheMainMenu,pressSystemConguration.
c.PressMore>More>ImportSetup.
d.HighlightthesystemsetupleyouwanttoimportfromtheSDcard.
e.PressImport.
f.Whenthefollowingmessageisdisplayed,pressOK:Congurationwas
successfullyimported
g.RemovetheSDcardandstoreitinasecurelocation.
10.ConguretheDeviceSettings:
a.OntheMainMenu,pressSystemConguration.
b.PressMore>More>ServiceSetup.
c.Whenprompted,entertheServicePassword.
Ifyoudonotknowthepassword,contactGEHealthcareTechnical
Support.