Maintenance
Thischapteraddressesmaintenancerecommendations,requiredtoolsandsupplies,
FRUreplacementprocedures,andfunctionalcheckoutprocedures.
RecommendedMaintenance
Regularmaintenance,irrespectiveofusage,isessentialtoensurethattheequipment
isalwaysfunctionalwhenrequired.
Refertotheoperator’smanualforyourdeviceforcleaningprocedures.Inaddition,
performtheprintheadcleaningprocedure,asdescribedin“CleaningthePrinthead”
onpage99.Thesystemdoesnotrequireanycalibrationprocedures.GEHealthcare
recommendsthatyouperformelectricalsafetychecksannually.Formoreinformation,
see“ElectricalSafetyChecks”onpage132.
WARNING:
EQUIPMENTFAILUREANDHEALTHHAZARDS—Failureonthepartofallresponsible
individuals,hospitalsorinstitutions,employingtheuseofthissystem,to
implementtherecommendedmaintenanceschedule,maycauseequipment
failureandpossiblehealthhazards.
Thesoleresponsibilityrestswiththeindividuals,hospitals,orinstitutionsutilizing
thesystem.
Themanufacturerdoesnot,inanymanner,assumetheresponsibilityfor
performingtherecommendedmaintenanceschedule,unlessanEquipment
MaintenanceAgreementexists.
RequiredToolsandSupplies
Thefollowingtoolsarerequiredtoperformtheproceduresdescribedinthischapter:
•ECGsimulator
•Smallatbladescrewdriver
•PosidriverPZ1x80
•PosidriverPZ2x100
•Currentleakagetester
•Anti-staticwriststrap
•Smallsidecutterpliers