ElectromagneticCompatibility(EMC)
EMCImmunityTest
ImmunityTest
Compliance
TestLevel
Compliance
Level
Electromagnetic
Environment–Guidance
Donotuseportableormobile
RFcommunicationsequipment
closertoanypartofthesystem,
includingthecables,than
therecommendedseparation
distancecalculatedforthe
equationapplicabletothe
frequencyofthetransmitter.
ConductedRF
3Vrms
150KHzto80
MHz
@2Hzmod.
MultilinkLead
set:3.0Vrms
ValueLeadset:
3.0Vrms
KISSLeadset:
0.3Vrms
Recommendedseparation
distance:
MultilinkLeadSet:
d=1.2√P
ValueLeadSet:
d=1.2√P
KISSLeadSet:
d=11.6√P
168
MAC™2000ECGAnalysisSystem
2053535-003C