7. Specifications
DMS 2 Operating Manual Page 175
7.2 Probe Specifications
7.2.1 Dual Element Probes
Nominal Contact Measurement Temp.
Model Probe Description Frequency Diameter Range (Steel) Range Cable
FH2E-D High Sensitivity Fingertip 8 MHz .38" .030" to 2.00" <130°F Potted
9.6 mm .75 to 50 mm <54°C
FH2E-D-REM High Sensitivity Fingertip 8 MHz .38" .030" to 2.00" <130°F Potted
with Remote Send Switch 9.6 mm .75 to 50 mm <54
o
C
KBA560-D General Purpose 5 MHz .625" .060" to 8.0" <250°F KBA-531
15.9 mm 1.5 to 200 mm <121
°C
HT400 High Temperature 5 MHz .500" .030" to 10.0" <1000°F KBA-535/536
HT400A 12.7 mm .75 to 250 mm <538°C
DM-401-GP General Purpose 5 MHz .700” 0.060” to 8.00 <300°F KBA-533
17.8 mm 1.5 to 200 mm <150°C
NOTES:
HT400A actual temperature range depends upon surface condition, and couplant.
HT400 high temperature probe is also compatible with a reduced measurement range.
HT400/400A temperature cycling required with surface temperatures above 400°F per Krautkramer
instruction card.