EasyManua.ls Logo

GE MFE DMS 2

GE MFE DMS 2
227 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4. Measuring Thickness
Page 130 DMS 2 Operating Manual
Step 1:
If necessary, activate the TG Primary Menu (TG
mode) by pressing
.
Step 2:
Activate the B-SCAN Submenu by selecting it with
or so that it’s highlighted. Five selections will appear
across the bottom of the display screen.
Step 3:
Press below the selection titled T-MIN. The
minimum thickness line can then be turned ON or OFF.
Step 4:
Turning the Minimum-Thickness graphic limit ON or
OFF will only display or remove the graphical “limit.” This action
will not otherwise effect the instrument’s alarm status.
Clearing the B-scan Display (B-SCAN–CLEAR)
The B-scan mode will allow you to map the thickness profile of
the material being tested over a period of time. This time
period may be set between 3.5 and 21 seconds. At the end of
the B-scan time period, the displayed B-scan will remain
displayed until it’s cleared with the CLEAR selection.
4.6 Saving, Recalling, Clearing TG Setups
(Parameter Sets)
The PARAM Submenu allows users to select, save and re-load
up to twenty thickness gage parameter sets with corresponding
A-scan images. When a parameter set is saved by selecting
SAVE P-SET, the settings of all of the parameters listed in
Table 4-1 are retained. This set of parameter settings (usually
associated with specific applications such as Tanks or Cooling
Tubes) can be applied to a data file by selecting LOAD P-SET.
When an existing P-SET is loaded or applied to a new data file,
the current instrument settings are automatically replaced by
the stored settings and the stored A-scan image is displayed.
These settings can then be adjusted as required.
The saved
P-SET can’t be changed
. If you wish to modify an existing
P-SET, first make the changes, delete the existing P-SET, as
described below, and save the new parameter settings using
the same P-SET name.
This feature provides the following two major benefits:
1. Storing and recalling TG calibration setups saves time,
ensures consistent data, and minimizes calibration errors.
Transferring setups to and from a computer also saves
time and allows identical setups to be stored in multiple
instruments.

Table of Contents

Related product manuals