Appendix C. Helpful Hints
Page 202 DMS 2 Operating Manual
is a 3-position buffer which cycles through the TG
Menu, File Navigation Mode, and DR Menu with successive
presses.
Set
’s function in the TG Secondary Menu
(I/O Submenu) to “DATREC” to store readings in on-board
recorder.
Set
’s function in the DR Menu (RECORD
Submenu) to THCK/THCK+ASCAN for storage of a
combined A-scan and thickness reading whenever
is pressed and held for two seconds.
Create a “Master Comment List” by pressing
BEFORE creating files in the File Creation Mode.
Comments (notes) can be attached to readings or
obstructs, but NOT to EMPTY LOCATIONS.
Pressing
within five seconds of storing a data point will
cause the just-stored point to become active and will open
the Active Comment List. Any comments selected will then
be attached to the just-stored point.
launches the File Creation Mode. It also creates a
Micro-Grid if pressed while an already-stored data point is
active.
Micro-Grids can only be attached to file locations that
contain thickness values (not empty or obstructed
locations)
Use the SAVE P-SET and LOAD P-SET selections in the
TG Primary Menu (PARAM Submenu) to name, store, and
recall parameter sets (i.e. single-element or frequently used
setups).
Appendix C: Helpful Hints