GE Analytical Instruments ©2006  13-12  DLM 14291 Rev. A 
•  Slowly unscrew the cap on the top of the purge vessel to release any 
pressure that might have built up during the purging. 
•  To drain the reducing agent, open the drain stopcock (CAUTION: HOT 
HYDROCHLORIC ACID). 
 
To refill the purge vessel with fresh reducing agent:  
 
•  Close the drain stopcock, add 4-5 mL of VCl
3
/HCl and 100 µL of dilute 
antifoaming agent. 
•  With the Outlet stopcock closed, replace the screw cap and open the gas 
inlet stopcock. Gas will begin to flow into the purge vessel 
•  Since the gas bubbler is still connected to the NOA, it is under reduced 
pressure. If the outlet stopcock on the purge vessel is opened rapidly, the 
NaOH will bump into the IFD filter. If the stopcock is opened slowly, the 
NaOH will not bump. 
•  Slowly open the outlet stopcok while watching the level of the NaOH.  
 
The NOA signal will initally increase with fresh VCl
3
/HCl, but the signal will 
return to baseline after a few minutes. 
 
Opening the Gas Bubbler 
 
The gas bubbler containing NaOH is also under a vacuum and if not opened 
properly, the NaOH can backup into the tubing to the purge vessel. To open the 
gas bubbler: 
 
•  Close the outlet stopcock on the purge vessel, leaving the stopcock on the 
bubbler open. 
•  Loosen the 1/4” nut on the Swagelok union connected to the outlet of the 
purge vessel. 
•  Once loose, slowly
 pull the union off the purge vessel while keeping an eye 
on the level of NaOH in the bubbler. As the union is removed, air will be