GE Analytical Instruments ©2006    DLM 14291 Rev. A 
6-2
A I N  
E N U         
                 
A N A L Y S I S C A L  I  B R A T  I O N
C O N T R O L    
E S S A G E S
A I N T E N A N C E       
0 4 J U L 0 0   1 2  :  3 4  :  5 6
 
C A L I B R A T  I O N       
                 
C A L I B R A T E         
V I  E 
           
P R I N T             
                 
 
L OG I N             
                 
O P E R A T O R         
S I E V E R S           
                 
                 
 
C A L I B R A T  I O N       
                 
Z E R O    G A S         
C A L I B    G A S         
                 
                 
 
                 
S u p p l y g a s    w  i  t  h    z  e  r o
N O c o n c e n  t  r  a  t  i  o  n   
                 
                 
E N T p r o c e e  d    C L R    e  s  c
 
                 
A I T t h e c  a  l  i  b  r  a  t  i  o n
i s b e i n g   p  e  r  f  o  r m e  d .
3 . 0 m V               
                 
C L R e  s  c  a p  e       
Calibration with Zero Air Filter 
 
To calibrate the zero offset using the filter, 
remove the Luer plug and brass Swagelok cap 
from the zero air filter. Connect the Luer 
fitting on the gas sampling line directly to the 
Luer adapter on the outlet of the filter and 
allow the NOA to draw ambient air through the 
filter for ~5 minutes.  
 
Calibrating with Zero Air Cylinder 
 
The setup for sampling gas from a pressurized 
gas source is shown in Figure 7-1. The Luer 
adapter tee included in the gas sampling 
package is connected to the outlet of the 
regulator on the gas cylinder, the PVC gas 
sampling line connected to one leg of the tee 
and the remaining leg is open to the 
atmosphere. The flow rate of gas from the 
pressurized source is adjusted to provide >200 
mL/min with the excess gas flowing out the 
open leg of the tee. Using this setup will ensure 
that the calibration is performed at the same 
flow rate as the measurements. Allow the NOA 
to sample the zero air for ~ 5 minutes.  
 
To calibrate of offset: 
 
•  Start the NOA and from the Measurement Menu, press CLEAR to switch to the 
Main Menu. 
 
•  Use the arrow buttons to scroll to Calibration and press ENTER. 
 
•  The Calibration Menu is displayed with three options: Calibrate, View and 
Print. 
 
•  Select Calibrate to display the Login menu.