GE Analytical Instruments ©2006    DLM 14291 Rev. A 
6-11
restrictors that provide a lower flow into the analyzer are available. Table I lists 
the restrictors available from Sievers Instruments. 
 
Table I. Available Flow Restrictors 
Part No.  Flow Rate (mL/min)*  Cell Pressure (torr)  Color Code 
HMI 14140  100  5.5  Black 
HMI 14120  55  4.0  Blue 
HMI 14110  30  3.2  Red 
HMI 14100  10  2.5  Green 
* - flow rates into NOA at atmospheric pressure of 630 torr 
 
To use the NOA at a lower flow rate, remove the standard flow restrictor. Install 
the new flow restrictor and recalibrate the NOA at the lower flow rate. 
 
Calibration at Lower Flow Rates 
 
Since the response of the NOA at flow rates below 200 mL/min will be lower than 
with the standard flow rate, the analyzer must be recalibrated for the lower flow 
rates. In most cases, high voltage for the PMT will need to be adjusted to achieve 
good results. Contact Sievers Instrument for the procedure to set the high voltage 
with the lower flow restrictors. If the high voltage is not adjusted, the NOA will 
report a response out of recommended limits” message during calibration.