GE Analytical Instruments ©2006  8-7  DLM 14291 Rev. A 
Sample collection using the Vital Capacity Bag kit consists of three steps: 
•  Connecting a pre-cleaned bag to the filler. 
•  Instructing the subject on how to perform the maneuver and collecting the 
samples 
•  Disconnecting the bag from the filler and sealing the bag. 
 
Connecting the bag to the filler 
 
To prepare the filler for sample collection:  
•  Stand the filler on the inspiratory filter. 
•  Remove the cap from the bag and grasp the cap, the top of the bag and 
internal tube with one hand while holding the bag filler with the other hand. 
•  Twist the bag slightly counterclockwise, then insert the Luer fitting on the bag 
into the Luer fitting on the filler and rotate fully clockwise until the o-ring 
seals against the base of the bag filler.  
•  Rotate the bag so that there are no folds or restrictions at the end of the tube. 
•  Open a new viral and bacterial filter and connect the filter to the tapered end 
on the top of the filler. 
•  Connect a sterile mouthpiece to the filter. 
 
Instructing the Subject and Collecting the Samples 
 
To perform the maneuver: 
•  Have the subject comfortably seated and holding the bag filler level. 
•  Have the subject place the mouthpiece in their mouth. 
•  Instruct the subject to inhale orally
 through the inspiratory filter to total lung 
capacity. 
•  Have the subject immediately exhale while watching the pressure gauge and 
adjusting the force of exhalation to maintain a pressure of 13 cm H
2
O to give 
an expiratory flow rate of 0.35 L/s.
* During the exhalation, the pressure should 
 
*  at 760 torr, for other pressures, calculate the required pressure from the equation: