GE Analytical Instruments ©2006  16-15  DLM 14291 Rev. A 
As standards and samples sit over time, NO from the air and nitrite and nitrate 
from the sample containers may leach into the solutions, resulting in increased 
peaks over time. This is particularly true for low level (< 1 µM) samples and 
standards. Thoroughly wash all containers with deionzed water and it may be 
necessary to prepare fresh standards, particularly at the lowest concentrations. 
 
High background Signal and Rising Baselines 
 
If there is a leak in the purge system, room air can be drawn into the NOA 
resulting in a high baseline signal. High baseline signals can also be observed if the 
purge vessel or reaction chamber is dirty, the purge gas is contaminated with NO 
or if the reagents are contaminated. These possible sources can be identified by 
recording the background without any reagents in the purge vessel or by trying 
another purge gas cylinder. 
 
The most common cause of increasing baseline signals over time is injection of 
materials that are slowly converted to nitric oxide such as L-NAME and other nitro-
containing compounds. If possible use other NOS inhibitors such as methyl arginine 
or other non-nitro-containing compounds (contact Sievers for additional 
information). 
 
Ghost Peaks 
 
When fresh reagent is added to the purge vessel, particularly the VCl
3
/HCl 
reagent, sometimes peaks are observed even when no injection is made. These are 
due to slow conversion of materials in the reagents to a species that react with 
ozone and will usually disappear as the reagent is purged. If ghost peaks are 
observed, continue purging the reagent until the baseline is stable. Extra peaks 
are also sometimes observed as the samples are being analyzed. These are due to 
compounds that are slowly converted to NO from previous injections. Purging the 
reagent (without injections) with usually eliminate these extra peaks.