GE Analytical Instruments ©2006  8-6  DLM 14291 Rev. A 
Figure 8-3: Setup for 
Cleaning Bags. 
and the final leg connected to the bag (Figure 8-
3). As outlined below, this setup allows zero air to 
be drawn into the bag. 
 
•  Remove the frit restrictor from the NOA and 
connect the Nafion drier directly to the NOA 
sample inlet. With 3-way valve and zero air 
filter connected to the NOA, turn the valve so 
that OFF is positioned toward the Zero Air 
Filter and allow the NOA to completely 
evacuate the gas from the bag. Disconnect the 
bag and repeat for the remaining bags. 
 
•  Connect the bag to a source of dry, low NO gas 
and slowly fill the bag. It is not necessary to completely fill the bag. 
 
•  Reconnect the bag to the 3-way valve. Turn the valve so that OFF points 
toward the zero air filter. Allow the NOA to completely evacuate the contents 
of the bag. 
 
•  After the bag is evacuated, turn the 3-way valve so that OFF points toward the 
sampling line connected to the NOA. This will allow gas from the zero air filter 
to be drawn into the Mylar bag and bring the pressure inside the bag back to 
atmospheric. Disconnect the bag, and seal the bag with the cap. Reposition the 
3-way valve so that OFF points toward the Zero Air Filter. Repeat for remaining 
bags. Store the bags in a clean, dry and secure location. Reinstall the 
restrictor before using the NOA for gas analysis. 
 
Bags cleaned by this procedure will remain low in NO for approximately 24 hours. 
 
Collecting the Samples – Vital Capacity Bag Kit