EasyManua.ls Logo

GE MOA 280i - Page 63

GE MOA 280i
211 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE Analytical Instruments ©2006 DLM 14291 Rev. A
6-3
A C C E P T N E
W
O F F S E T S
O L D P P
M
0 . 0
N E
W
P P
M
0 . 0 .
O L D P P B 1 0 . 3
N E
W
P P B 1 0 . 6
E N T a c c e p t C L R r e c e c t
A C C E P T N E
W
O F F S E T S
O L D P P
M
0 . 0
N E
W
P P
M
0 . 0 .
O L D P P B 1 0 . 6
N E
W
P P B 9 . 7
E N T a c c e p t C L R r e j e c t
C a l i b r a t i o n s a m p l e
h a s n o t b e e n s t a b l e .
E N T E R \ C R L t o c o n t i n u e
B o t h o f f s e t s e x c e e d
r e c o m m e d l i m i t s .
E N T p r o c e e d C L R r e j e c t
A C C E P T N E
W
O F F S E T S
O L D P P
M
0 . 0
N E
W
P P
M
5 . 2 .
O L D P P B 1 0 . 3
N E
W
P P B 1 0 9 . 5
E N T a c c e p t C L R r e j e c t
If the security is not enabled, then the menu
will display Operator and Sievers options. If the
security features are enabled, a list of the
administrators is displayed and a password is
required to calibrate the NOA (see Chapter 14).
Select Operator to display the Calibrate
Menu.
Select the Zero Gas option.
The zero gas calibration instruction menu is
displayed, after the zero gas has been
flowing into the NOA for ~5 minutes, press
ENTER to start the zero gas calibration.
The analyzer switches displays and shows a
moving average of the PMT signal in millivolts
with the amplifier in the low sensitivity range.
The moving average starts at 0 mV and slowly
increases. To ensure a accurate calibration, the
firmware compares the most recent measurement
of the ozone background signal to the moving
average until a stable measurement is obtained.
The analyzer then switches to the high sensitivity setting, resets the moving average to
0 mV and measures background signal until a stable signal is obtained. The ppb and
ppm offsets are then calculated and displayed along with the previous calibration. The
units for the offsets are counts at the analog-to-digital converter, with each count
being ~ 0.2 ppb or 0.1 ppm.
Press ENTER to accept the new calibration values and return to the Calibrate Menu.
Press Clear to reject the new calibration.
Two or three successive Zero Gas calibrations should be performed to make sure
that the calibration is not drifting up or down. Replicate zero gas calibrations
should agree within 1 or 2 counts. If the calibrations are drifting, continue sampling
the zero air until stable replicate calibrations are obtained.

Table of Contents

Related product manuals