EasyManua.ls Logo

GE MOA 280i - Page 174

GE MOA 280i
211 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE Analytical Instruments ©2006 15-6 DLM 14291 Rev. A
Figure 15-2: Schematic of NOA-Procedure for Cleaning Reaction Cell.
Turn off the main power switch and remove the AC power cord from the power
entry module.
Allow the analyzer to stand for a couple of hours (overnight is best) so the PMT can
warm to room temperature.
Make sure the power cord is disconnected, and then remove the NOA cover by
removing the phillips-head screws on the side of the analyzer.
Carefully lift the cover off the NOA.
Remove the two coaxial cables connected to the PMT socket.
In order to minimize exposure of the PMT to light, turn off the room lights
before removing the PMT.
Unscrew the PMT socket counterclockwise until the socket is clear of the
cooler.
Remove the PMT from the cooler, but leave the PMT in the socket.

Table of Contents

Related product manuals