GE Analytical Instruments ©2006 DLM 14291 Rev. A
2-2
Power Requirements................................120 V, 60 Hz (6A)
100 V, 50 or 60 Hz (7A)
230 V, 50 Hz (3A)
NOA
.........................................................Height 16 in. (41 cm)
.........................................................Width 6.2 in. (16 cm)
.........................................................Length 20 in. (51
cm) ....................................................
.........................................................Weight 35 lbs. (16
kg)
Vacuum Pump with installed trap
.........................................................Height 14.5 in. (37 cm)
.........................................................Width 7.5 in. (19 cm)
.........................................................Length 19 in. (48
cm)
.........................................................Weight 47 lbs. (21.5
kg)
Operating Environment
Ambient Temperature ....................................32°F to 86°F (0°C to 30°C)
Relative Humidity.......................................... 0% to 90%
Fuse Requirements
Main Fuse ...................................100 VAC model: T, 10A, 250V
120 VAC model: T, 5A, 250V
230 VAC model: T, 5A, 250V
Ozone Fuse..................................100VAC model: T, 200mA, 250V
120 VAC model: T, 200mA, 250V
230 VAC model: T, 100mA, 250V