GE Analytical Instruments ©2006 8-4 DLM 14291 Rev. A
Figure 8- 2: Diagram of Deadspace Discard Bag Collection Kit.
Screw the pressure gauge into the bag filler body (Teflon tape is not required).
Inspect the pressure gauge. If the needle does not read 0 cm H
2
O, unscrew the
clear plastic cover of the gauge and adjust the screw at the base of the gauge
with a screwdriver until the needle points to 0 cm H
2
O. Replace the clear
plastic cover of the gauge.