Chapter 4. Programming
74 TransPort
®
PT900 User’s Manual
4.6.2a Custom Transducer Spacing
If your transducers were installed with a spacing different from the value calculated
by the APP, refer to Figure 81 below and input the actual spacing as follows:
Note: Be sure that, if the installed spacing you input is larger than the calculated
value, it is not more than 10% above the calculated value.
1. Enter your actual transducer spacing in the units specified by your system
UNITS setting.
2. Click the
OK button to accept the new value.
Figure 81: Custom Transducer Spacing