EasyManua.ls Logo

General Monitors 4802A - Calibration Check Mode; Figure 39 - Portable Purge Calibrator; Figure 40 - Entering CAL Check Mode

Default Icon
75 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 4802A
20
5.2 Calibration Check Mode
To perform a calibration check, follow the procedure listed below.
1. Place the cup from the portable purge calibrator over the sensor (Figure 39)
NOTE: Do not attach the cup from the portable purge calibrator to the sensor,
until display shows a flashing Zero.
Figure 39 – Portable Purge Calibrator
* The Calibration Check Mode cannot be entered if the 4802A is in alarm.
2. Enter the Calibration Check mode by pressing and holding the Mode/Select switch
until the CAL LED begins to flash (about ten seconds). When the CAL LED begins
to flash, release the Mode/Select switch. The unit is now in the Calibration Check
Mode (Figure 40).
Figure 40 – Entering CAL Check Mode

Table of Contents

Related product manuals