Voltage
(V)
Connection Frequency
(Hz)
Power
(kW)
Pre
fuse (A)
(Curve C)
Connection cable
Area
(mm2)
Length
(m)
Color code
240 3+PE 50 7.0 3 x 20 4 x 4.0 3.0 L
1
=Black
L
2
=Brown
L
3
=Gray
PE=Green/
Yellow
230 3+PE 60 7.0 3 x 20 4 x 4.0 3.0 L
1
=Black
L
2
=Brown
L
3
=Gray
PE=Green/
Yellow
230 3+PE 50 7.0 3 x 20 4 x 4.0 3.0 L
1
=Black
L
2
=Brown
L
3
=Gray
PE=Green/
Yellow
230 1N+PE 50 2.2 1 x 10 3 x 2.5 3.0 L
1
=Black
N=Blue
PE=Green/
Yellow
220 3+PE 60 7.0 3 x 20 4 x 4.0 3.0 L
1
=Black
L
2
=Brown
L
3
=Gray
PE=Green/
Yellow
220 3+PE 50 7.0 3 x 20 4 x 4.0 3.0 L
1
=Black
L
2
=Brown
L
3
=Gray
PE=Green/
Yellow
2 INSTALLATION
18 (48) INSTALLATION MANUAL GETINGE WD15 CLARO