Voltage
(V)
Connection Frequency
(Hz)
Power
(kW)
Pre
fuse (A)
(Curve C)
Connection cable
Area
(mm2)
Length
(m)
Color code
208 3+PE 60 6.3 3 x 20 4 x 12
(AWG)
3.0 L
1
=Red
L
2
=White
L
3
=Black
PE=Green/
Yellow
200 3+PE 60 5.9 3 x 20 4 x 4.0 3.0 L
1
=Black
L
2
=Brown
L
3
=Gray
PE=Green/
Yellow
200 3+PE 50 5.9 3 x 20 4 x 4.0 3.0 L
1
=Black
L
2
=Brown
L
3
=Gray
PE=Green/
Yellow
Voltage variations allowed:
short term from nominal voltage +- 10%.
2 INSTALLATION
INSTALLATION MANUAL GETINGE WD15 CLARO 19 (48)
<Doc_INS><Doc_6001396702><Rev.I><Lang_en>