Inspections
Electrical inspections
6
Maquet PowerLEDII
01810 EN 03
43 / 56
XHD1 monitor mount
or
1
2
3
Fig.56: XHD1 electrical safety test
If the monitor mount does not include a handle option, the measurement should be made
between the flange 1 and the earth wire connection on the VESA block 3 .
If the monitor mount includes a handle option, the measurement should be made between
the flange 1 and the earth wire connection on the handle 2 .