EasyManuals Logo

Getinge Maquet PowerLED II Maintenance Manual

Getinge Maquet PowerLED II
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #42 background imageLoading...
Page #42 background image
6
Inspections
Electrical inspections
42 / 56
Maquet PowerLEDII
01810 EN 03
XHS0 monitor mount
or
1
2
3
4
Fig.55: XHS0 electrical safety test
If the monitor mount does not include any accessories, the measurement should be made
between the flange 1 and one of the monitor mounting screws 2 .
If the monitor mount includes a handle option, the measurement should be made between
the flange 1 and one of the handle mount mounting screws 3 . Once the measurement is
complete, replace the cap.
If the monitor mount includes a Rear Box option, the measurement should be made between
the flange 1 and one of the case mounting screws 4 , with case cover removed. Once the
measurement is complete, replace the cover.

Other manuals for Getinge Maquet PowerLED II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet PowerLED II and is the answer not in the manual?

Getinge Maquet PowerLED II Specifications

General IconGeneral
BrandGetinge
ModelMaquet PowerLED II
CategoryMedical Equipment
LanguageEnglish