EasyManua.ls Logo

GRAPHTEC GL900 - Input Unit Specifications

GRAPHTEC GL900
97 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-4
Specifications
Input Unit Specifications
Type Measurement temperature range Measurement accuracy
R/S 0 TS 100 ±7.0°C
100 < TS 300 ±5.0°C
R:300 < TS 1600 ±(0.05% of rdg +3.0°C)
S:300 < TS 1760 ±(0.05% of rdg +3.0°C)
B 400 TS 600 ±5.5°C
600 < TS 1820 ±(0.05% of rdg +3.0°C)
K -200 TS -100 ±(0.05% of rdg +3.0°C)
-100 < TS 1370 ±(0.05% of rdg +2.0°C)
E -200 TS -100 ±(0.05% of rdg +3.0°C)
-100 < TS 800 ±(0.05% of rdg +2.0°C)
T -200 TS -100 ±(0.1% of rdg +2.5°C)
-100 < TS 400 ±(0.1% of rdg +1.5°C)
J -200 TS -100 ±3.7°C
-100 < TS 100 ±2.7°C
100 < TS 1100 ±(0.05% of rdg +2.0°C)
N0 TS 1300 ±(0.1% of rdg +2.0°C)
W0 TS 2315 ±(0.1% of rdg +2.5°C)
Reference contact compensation accuracy ±1.0°C
*1: Thermocouple diameters T: 0.32
φ
, others: 0.65
φ
Item Description
Number of input channels Fixed to 8 channels
Input terminal Voltage BNC connector
type
Temperature
M3 screw type terminal board
Input method All channels isolated
Imbalanced input
Simultaneous sampling of all channels
Maximum sampling speed 10 μs
Measurement Voltage 20 m, 50 m, 100 m, 200 m, 500 m;
ranges 1, 2, 5, 10, 20, 50, 100, 200, 500 V; F.S., 1-5 V F.S.
Temperature
Thermocouples : K, J, E, T, R, S, B, N, W (WRe5-26)
Humidity 0 to 100% (voltage 0 V to 1 V scaling conversion) * with B-530 (option)
Measurement accuracy*
1
Voltage
(23°C±5°C) ±0.25% of F.S.
When 30 minutes or more Thermocouple
have elapsed after power
was switched on
Filter Line
GND connected
Standing upright
When the measured value
is Average
Reference contact Internal/External switching
compensation accuracy
A/D converter 16 bits (out of which 14 bits are internally acknowledged)
Temperature coefficient Gain : 0.01% of F.S./°C
Zero : 0.02% of F.S./°C
Input resistance 1 MΩ±5%
Allowable signal source Within 1 kW
resistance
Maximum permissible Between input channel + and - terminals : 20 mv to 1 V 30 Vp-p
input voltage : 2 V to 500 V 500 Vp-p
Between input channel terminals : 60 Vp-p
Between input channel terminal and GND terminal : 60 Vp-p
Withstand voltage Between input channel terminal and GND terminal : 1 minute at 1000 Vp-p
Between input channel terminals : 1 minute at 1000 Vp-p
Insulation resistance Between input channel terminal and GND terminal : At least 50 MΩ (at 500 VDC)
Common mode rejection ratio
At least 90 dB (50/60 Hz; signal source 300 Ω or less)
S/N (Noise) 20 mV range : At least -40 dB
Other range : At least -50 dB
Frequency response DC to 20 KHz (+1/-4 dB)
Filter OFF, Line, 5Hz, 50Hz, 500Hz
(Attenuation) -3 dB / 6 dB oct

Table of Contents