22.7 SCANNER ALWAYS CONNECTS TO AN UNWANTED WLAN AND DOES NOT GO INTO WLAN ACCESS
POINT MODE
Solution: Power on the scanner and wait at least 30 seconds for the scanner to start. Press the power button
for 2 seconds – release the finger when the light blinks blue. Now the WLAN access point is activated and can
be connected to. Note: Pressing the power button for too short (less than 2 seconds) or for too long time
(more than 8 seconds) initiates scanner shutdown.
22.8 SCANNER WLAN ACCESS POINT PASSWORD IS FORGOTTEN
Solution: Power on the scanner and wait at least 30 seconds for the scanner to start. Press the power button
for 2 seconds – release the finger when the light blinks blue. Now the WLAN access point is activated. Press
the power button again for 2 seconds – release the finger when the light blinks cyan. The WLAN access point
password is reset back to the same as in device label.
22.9 DOWNLOAD DOES NOT START AUTOMATICALLY WHEN EXPORTING TO WebDL
Solution: Check your web browser settings and disable all ad-blockers and pop-up blockers for
*.grundium.net.
23 CERTIFICATIONS
CE statement / EU Declaration of
Conformity
Grundium Ltd. hereby declares that this
device is in compliance with the essential
requirements and other relevant provisions
of EU regulation 2017/746.
Grundium Ocus® is
an In Vitro
diagnostic Medical
Device.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) This device must accept any interference received, including interference that may cause undesired operation.
Any change or modification to the product not expressly approved by Grundium Oy could void the user's authority to
operate the device.
This equipment complies with FCC RF radiation exposure limits set forth for an uncontrolled environment.
This device has been SAR tested for FCC body and extremity limits. The maximum SAR value reported is 0.25W/kg.