Weighing with a Class III scale
Ta r e
Scale W1
0.0 kg
NET
Exit
Ta r e
80.3 kg
NET
Exit
1. Always reset the scale module before weighing. See section ”Resetting of
Class III scale”.
2. Lift the lifting sling o the hanger. Place the patient in the lifting sling before
mounting it on the lifting hanger.
3. Lift the patient. The correct weight can be read when ”kg” appears in the
display. When ”kg” is permanently shown in the display, the scale is at rest
and can be read. When ”kg” disappears, the scale is no longer at rest and the
weight displayed may be incorrect.
While being weighed, the user must be free of the surroundings in order to
not aect the weighing. Do not pull the hand control while weighing, as doing
so will aect the result.