33
© Guldmann GB/US-02/2021 • # 550216_19
© Guldmann GB/US-02/2021 • # 550216_19
Scale Info 2-2
Changes
001
Next Exit
Changes (information screen 2 of 2 in scale menu): change counter
Counts changes in software parameters such as calibration and
values for G factor
Error notications
- - -
Scale W2
Exit
The scale is overloaded.
Please follow the scale’s prescribed load (see identication label on
the hoist).
- - -
Scale W1
Reset Exit
The scale is below the minimum possible reading value
(in other words, the load weighed is under 2 kg).
Select ”Reset” and start over with setting the scale.
004
Scale
ExitReset
An error has occured in the scale. The display shows an error code.
Call a technichian to help solve the problem.