COMMAND OVERVIEW
123
Measure Commands
MEASure:VOLTage:DC?
Returns the DC voltage measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:VOLT:DC?
>+0.488E-4
Returns the DC voltage measurement as 0.0488 mV.
MEASure:VOLTage:AC?
Returns the AC voltage measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:VOLT:AC?
>+0.511E-3
Returns the AC voltage measurement as 0.511 mV.
MEASure:VOLTage:DCAC?
Returns the DC+AC voltage measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:VOLT:DCAC?
>+0.326E-3
Returns the DC+AC voltage measurement as 0.326 mV.
MEASure:CURRent:DC?
Returns the DC current measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:CURR:DC?
>+0.234E-4
Returns the DC current measurement as 0.0234 mA.