EasyManua.ls Logo

GW Instek GDM-834X Series - Sense Commands; [Sense:]Detector:rate; [Sense:]Temperature:rjunction:simulated; [Sense:]Temperature:tcouple:type

GW Instek GDM-834X Series
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
COMMAND OVERVIEW
127
SENSe Commands
[SENSe:]TEMPerature:TCOuple:TYPE
Sets thermocouple type.
Parameter: Type(J | K | T)
Example: SENS:TEMP:TCO:TYPE J
Sets the thermocouple to type J.
[SENSe:]TEMPerature:TCOuple:TYPE?
Returns the thermocouple type.
Return parameter: J, K, T
[SENSe:]TEMPerature:RJUNction:SIMulated
Set temperature simulation value.
Parameter: <NRf>(0.00 ~ 50.00)
Example: SENS:TEMP:RJUN:SIM 25.00
Sets the thermocouple junction temperature to 25˚C.
[SENSe:]TEMPerature:RJUNction:SIMulated?
Returns temperature simulation value.
Return parameter: <NR1> (+0000~+5000) ,where +0000=0.00˚C,
+5000=50.00˚C
[SENSe:]DETector:RATE
Sets the detection rate (sample rate)
Parameter: RATE(S | M | F)
Example: SENS:DET:RATE S
Sets the rate to slow (S).
[SENSe:]DETector:RATE?
Returns the sample rate.
Return parameter: SLOW, MID, FAST