1st Unit
The units for the reading on the
primary display.
2nd Value
The reading on the secondary
display.
2nd Unit
The units for the reading on the
secondary display.
Count
Counts the number of readings
each time the measurement is
started. The count is restarted
each time measurement is
restarted. When a measurement
is started/restarted, the first
count is marked as #START#,
the last as #END#.
Note
Records the accumulative
number of readings that are
recorded in that file, up to the
maximum of 50,000.
Example:
Time(dd)
Time
(hh:mm:ss)
1st Value
1st Unit
2nd
Value
2nd
Unit
Count
Note
0
0:00:05
0.00E+00
V DC
--
--
#START#
00001#
0
0:00:06
0.00E+00
V DC
--
--
2
00002#
0
0:00:06
0.00E+00
V DC
--
--
#END#
00003#
Filename Format
Overview
When files are saved to USB they are saved as a
number starting from GW000\GW000-XX.CSV
and are automatically incremented for each
new CSV file*. For example: the first file will be
named, GW000\GW000-XX.CSV, the next
GW001\GW001-XX.CSV and so on.
Note that the suffix, XX, represents a number
from 00 to 99. Each time the system logs more
than 50000 readings in total*, a new file is
generated and the suffix is incremented. For