GDM-834X Series User Manual
36
For example:
A
+5V
0V
-5V
C D
B
E
DC 5V Range
A,B: Input exceeds the ADC dynamic range.
C,D: The DCV offset causes the input to exceed the
upper ADC dynamic range.
E: The DCV offset causes the input to exceed the
lower ADC dynamic range.
The DC voltage range should be manually selected
when any of the following conditions are true:
When DCV measurement is used.
When the signals being measured contain both
DC and AC components.
When the amplitude of the AC component in
the measured signal is higher or lower than the
dynamic range of the range being currently
selected by the auto-range function.
Selected DCV Range
ADC Dynamic Range
DC 500mV
max±100mV
DC 5V
max±1V
DC 50V
max±10V
DC 500V
max±100V
DC 1000V
max±1000V