GDM-906X Series User Manual
282
CONFigure:CAPacitance
Sets measurement to Capcitance on the 1
st
display and specifies range.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF),]
Example: CONF:CAP 10e-7
Sets the Capcitance range to 100nF.
CONFigure:CONTinuity
Sets measurement to Continuity on the 1
st
display.
Parameter: [None]
CONFigure:DIODe
Sets measurement to Diode on the 1
st
display.
Parameter: [None]
CONFigure:TEMPerature
Sets measurement to Temperature on the 1
st
display and specifies
type/resolution.
Parameter: [None] | [Probe type [, Type [, 1 [, Resolution(<NRf> | MIN | MAX
| DEF)]]]]
<Probe type>:TCOuple, RTD, FRTD, THERmistor, FTHermistor.
<Type>:
Tcouple: J | K | N | R | S | T | B | E | USER
RTD / FRTD : PT100 | D100 | F100 | PT385 | PT3916 | USER
Thermistor / Fthermistor : 2.2k | 5k | 10k | USER
Example: CONF:TEMP TCO,K
Sets the measurement mode to TCO with a type K sensor.
Autoranging (AUTO or DEFault), will generate an error if you specify a
<resolution> because the instrument cannot accurately resolve the
integration time (especially if the input continuously changes). If your
application requires autoranging, specify DEFault for the <resolution> or
omit the <resolution> altogether.
Secondary Display: CONFigure2 Commands