REMOTE CONTROL
283
CONFigure2[:VOLTage]:DC
Sets measurement to DC Voltage on the 2
nd
display and specifies
range/resolution.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX |
DEF)[,Resolution(<NRf> | MIN | MAX | DEF)]]
Example: CONF2:VOLT:DC 1,MAX
Sets the voltage range to 1 volt and the resolution to the maximum.
Autoranging (AUTO or DEFault), will generate an error if you specify a
<resolution> because the instrument cannot accurately resolve the
integration time (especially if the input continuously changes). If your
application requires autoranging, specify DEFault for the <resolution> or
omit the <resolution> altogether.
CONFigure2[:VOLTage]:AC
Sets measurement to AC Voltage on the 2
nd
display and specifies range.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF)]
Example: CONF2:VOLT:AC
Sets the measurement mode to AC voltage on the 2
nd
display.
CONFigure2:CURRent[:DC]
Sets measurement to DC Current on the 2
nd
display and specifies
range/resolution.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX |
DEF)[,Resolution(<NRf> | MIN | MAX | DEF)]]
Example: CONF2:CURR:DC 10e-3,DEF
Sets the DC current range to 10mA using the default resolution on the 2
nd
display.
Autoranging (AUTO or DEFault), will generate an error if you specify a
<resolution> because the instrument cannot accurately resolve the
integration time (especially if the input continuously changes). If your
application requires autoranging, specify DEFault for the <resolution> or
omit the <resolution> altogether.
CONFigure2:CURRent:AC
Sets measurement to AC Current on the 2
nd
display and specifies range.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF)]
Example: CONF2:CURR:AC 10e-2
Sets the measurement mode to AC Current with a 100mA range on the 2
nd
display.
CONFigure2:FREQuency
Sets measurement to Frequency on the 2
nd
display and specifies range.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX |
DEF)[,Resolution(<NRf> | MIN | MAX | DEF)]]
Example: CONF2:FREQ MAX
Sets the frequency range to max on the 2
nd
display.