COMMAND OVERVIEW
117
CALC:SCAN:CHAN 5
Sets the channel to 5.
Sets or returns the interval delay for the scan function.
CALCulate:SCAN:DELay <NR1>
CALCulate:SCAN:DELay?
Parameter/
Return parameter
CALC:SCAN:DEL 500
Sets interval delay of the scan to 500ms.
CALCulate:SCAN:LIMit:LOWer
Sets or returns the lower limit value for the scan function.
CALCulate:SCAN:LIMit:LOWer {<NRf>[,<String>]}
CALCulate:SCAN:LIMit:LOWer?
mohm/ohm/kohm/maohm,unit
If unit is not set, the unit will be
automatically set by the present range.
CALC:SCAN:LIM:LOW 0.123,maohm
Sets the lower limit value to 0.123MΩ.
CALC:SCAN:LIM:LOW?
>0.1230E+6
Returns the lower limit as 0.123MΩ.
CALCulate:SCAN:LIMit:MODE
Sets or returns the scan function compare mode.
CALCulate:SCAN:LIMit:MODE {ABS|DPER}
CALCulate:SCAN:LIMit:MODE?
Parameter/
Return parameter
The test results are judged from
absolute values.