GOM-804 & GOM-805 User Manual 
118 
The test results are judged from a 
reference value ±  a percentage offset. 
(delta percent) 
CALC:SCAN:LIM:MODE ABS 
Sets compare mode to absolute values. 
CALCulate:SCAN:LIMit:REFerence 
 
Sets or returns the reference limit for the scan function. 
CALCulate:SCAN:LIMit:REFerence {<NRf>[,<String>]} 
CALCulate:SCAN:LIMit:REFerence? 
mohm/ohm/kohm/maohm,unit 
If  unit is not set,the unit will be 
automatically set by the present range. 
CALC:SCAN:LIM:REF 10.00,mohm 
Sets the reference limit to 10.00mΩ. 
CALC:SCAN:LIM:REF? 
>10.0000E-3 
Returns the reference limit as 10.00mΩ. 
CALCulate:SCAN:LIMit:UPPer 
 
Sets or returns upper limit of  the scan function. 
CALCulate:SCAN:LIMit:UPPer {<NRf>[,<String>]} 
CALCulate:SCAN:LIMit:UPPer? 
mohm/ohm/kohm/maohm,unit 
If  unit is not set, the unit will be 
automatically set by the present range. 
CALC:SCAN:LIM:UPP 1.37,kohm 
Sets the upper limit to 1.37kΩ. 
CALC:SCAN:LIM:UPP? 
>1.3700E+3 
Returns the upper limit as 1.37kΩ. 