GOM-804 & GOM-805 User Manual
124
SENSe:REL:DATa
Set
Query
Description
Sets or returns the relative value for the relative function.
Syntax
Query Syntax
SENSe:REL:DATa <NRf>
SENSe:REL:DATa?
Parameter
<NRf>
0.0000~500.00
The unit will be auto set by the present
range.
Return parameter
<NR3>
± 0.0000~5.1000E± X
Example
SENS:REL:DAT 490.32
Sets the relative function value to 490.32Ω.
SENS:REL:DAT?
>4.9032E+2
Returns the relative value (490.32Ω).
SENSe:REL:STATe
Set
Query
Description
Sets or returns the relative function state.
Syntax
Query Syntax
SENSe:REL:STATe <NR1> | {OFF|ON}
SENSe:REL:STATe?
Parameter/
Return parameter
<NR1>
0:OFF.
1:ON.
OFF
Turn the relative function off.
ON
Turn the relative function on.
Example
SENS:REL:STAT OFF
Sets the relative function off.
SENSe:REALtime:STATe
Set
Query
Description
Sets or returns the real time function state.
Syntax
Query Syntax
SENSe:REALtime:STATe <NR1> | {OFF|ON}
SENSe:REALtime:STATe?