COMMAND OVERVIEW
135
TEMPerature:CONVersion:DISPlay
Sets or returns the temperature display mode for the
temperature conversion function.
TEMPerature:CONVersion:DISPlay <NR1>
TEMPerature:CONVersion:DISPlay?
Parameter/
Return parameter
TEMP:CONV:DISP 1
Sets the temperature display mode for the temperature
conversion function is ΔT.
TEMPerature:CONVersion:MATH:DATa
Returns conversion function deviation value.
TEMPerature:CONVersion:MATH:DATa?
TEMP:CONV:MATH:DAT?
Returns 1.250E+2.
TEMPerature:CONVersion:RESistance
Sets or returns the initial resistance for the temperature
conversion function.
TEMPerature:CONVersion:RESistance {<NRf>[,<String>]}
TEMPerature:CONVersion:RESistance?
mohm/ohm/kohm/maohm,unit
If the unit is not set, the unit will be
automatically set by the present range.
www.GlobalTestSupply.com