The Scan function sequentially scans up to 100 channels 
and grades the resistance of  the DUT on each channel to 
a reference value. An automated handler or test fixture is 
required to interface the DUTs to the measurement 
terminals and the scan interface that controls the timing 
of  each scan.   
 
Note: The automated handler/test fixture is 
user-supplied. Please see your distributor for support and 
technical details.   
Grading of  each DUT is essentially the same as the 
compare function (page 41), the difference being the 
Scan function will compare up to 100 DUTs sequentially, 
whereas the Compare function will compare only one 
DUT at a time. 
The scan function compares a measured value to a 
“Reference” value that has an upper (HI) and lower (LO) 
limit. If  the measured value is within the upper and lower 
limit, then the measured value is judged as IN.     
There are two modes that can be used to make a 
judgment: ABS and  △% modes.   
The ABS mode compares the measured value to the 
upper (HI) and lower (LO) limits. The upper and lower 
limits are set as absolute resistance values.   
The  △% compare function compares the deviation of