APPENDIX 
    153 
temperature and the calibration temperature. 
 
“------” indicates over range 
20 sets of comparator status can be selected. 
Dry Resistance Measurement 
 
Temperature Measurement 
 
Temperature sensor 
(option) 
Platinum resistor. 
Lead length: 1.5m approx. 
Temperature Correction Function 
 
Reference temperature 
range 
Thermal coefficient 
range 
Accuracy of temperature compensation for 3930 
ppm/Cu wire.* 
0.3%+resistance measurement accuracy. 
0.6%+resistance measurement accuracy. 
*The temperature coefficient for the other settings must be calculated 
individually according to different conditions. 
*If the temperature coefficient or the difference between the environmental 
temperature and the required temperature exceeds normal operation, after 
calculating the compensation, the variation to the reading value will be 
significant. 
*When using the PT-100 temperature sensor for temperature measurements, 
the accuracy of the sensor (typical accuracy of <±0.5ºC) should also be taken 
into account and calculated for. 
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com