GOM-804 & GOM-805 User Manual
80
Pin Definitions for the Handler
Interface
As this interface is used for the handler and scan functions, the interface
pinout depends on the function mode. The following pinout is only applicable
when using the Binning or Compare function.
HANDLER / SCAN / EXT I/O
1
14
13
25
Handler Interface for Binning and Compare Functions
Trigger for a single measurement.
High when the compare result is
either HI or LO (fail).
High when the compare result is
deemed HI.
High when the compare result is IN
(pass).
High when the AD conversion has
completed. The DUT is ready to be
changed.
High when the binning sorting result
is within the bin1 setting range.
High when the binning sorting result
is within the bin2 setting range.
High when the binning sorting result
is within the bin3 setting range.
High when the binning sorting result
is within the bin4 setting range.