Example—Calculating the resistance of  a PT-100 
RTD at 100˚C (T). The following R
0
 (Ωat 0˚C), 
alpha, beta, and delta values are used for the   
PT-100 RTD: 
  T=100˚C 
  R0 (Ω at 0˚C) = 100Ω 
  Alpha=0.003850 
  Beta=0.10863 
  Delta=1.49990 
A, B, and C are calculated according to equations 
listed above: 
  A=0.00391 
  B=5.77e-7 
  C=4.18e-12 
The resistance of  the RTD at 100˚C (R
100
) is then 
calculated as follows: 
R100:  =R0[1+AT=BT
2
+CT
3
(T-100)] 
  =100{1+[(0.00391)(100)]+[(-5.77e-7)(100
2
)
  +[(-4.18E-12)(100
3
)(100-100)]]} 
  =138.5Ω