GOM-804 & GOM-805 User Manual
112
Example
BINN:LIM:DISP DPER
Sets the mode toΔ %.
BINNing:LIMit:REFerence
Set
Query
Description
Sets or returns the limit reference value for the bin
sorting function.
Syntax
Query Syntax
BINNing<X>:LIMit:REFerence {<NRf>[,<String>]}
BINNing<X>:LIMit:REFerence?
Parameter
<NRf>
000.0001~999.9999
<String>
mohm/ohm/kohm/maohm,unit
If the unit is not set, the unit will be
automatically set by the present range.
Return parameter
<NR3>
000.0001~999.9999E± X
Example
BINN:LIM:REF 100
Sets the limit reference value to 100Ω.
BINN:LIM:REF?
>100.0000E+0
Returns the reference as 100Ω.
BINNing:LIMit:RESult
Query
Description
Returns the bin sorting function test result.
Query Syntax
BINNing:LIMit:RESult?
Return parameter
<NR1>
1~8: Bin1~Bin8
9: Bin Out
Example
BINN:LIMit:RES?
>1
Indicates a pass for bin1.