COMMAND OVERVIEW
133
Temperature Commands
TEMPerature:AMBient:DATa
Set
Query
Description
Sets or returns the user-set ambient temperature value for
the temperature compensation and the temperature
conversion function.
Syntax
Query Syntax
TEMPerature:AMBient:DATa <NRf>
TEMPerature:AMBient:DATa?
Parameter
<NRf>
-50.0~399.9 (Unit: ºC)
Return parameter
<NR2>
-50.0~399.9 (Unit: ºC)
Example
TEMP:AMB:DAT 25.6
Sets the user ambient temperature value to +25.C.
TEMP:AMB:DAT?
>25.6
Returns the set ambient temperature as 25.6ºC .
TEMPerature:AMBient:STATe
Set
Query
Description
Sets or returns the state of the user-set ambient
temperature.
Syntax
Query Syntax
TEMPerature:AMBient:STATe <NR1> | {OFF|ON}
TEMPerature:AMBient:STATe?
Parameter/
Return parameter
<NR1>
0:OFF.
1:ON.
OFF
Disables the user-set ambient
temperature.
ON
Enables the user-set ambient
temperature.
Example
TEMP:AMB:STAT OFF
Disables the user-set ambient temperature.